Hidex ULLA
Hidex has served the liquid scintillation counting community for three decades. In recent years, the global market has been missing a true ultra low level LSC that would meet the most demanding needs. Applications in hydrogeology when mapping the global clean water resources, detection of biogenic carbon content in materials or measuring trace concentrations of alpha and beta isotopes in soil, food and drinking water require absolute accuracy and precision. Hidex ULLA combines existing technology, latest innovation and modern usability.
Key Features
Ultra Low Background and Triple Coincidence Detection
High Counting Efficiency
Hidex ULLA:
Unquenched: >70%
3H in water: >35%
High quench: 10%
Quench Correction with TDCR
Luminescence-Free Counting
Alpha/Beta Separation
Ideal Applications
C-14 dating
Biogenic carbon detection in fuels
Alpha and beta isotopes in soil, food, and water
Hidex VALO Software
Technical Specifications
Sample capacity, 20/7/5mL
40/96
Counting efficiency H-3/C-14 (%)
70/97 (unquenched)
> 35 for ³H in 8+12 ml H₂O
Background H-3 in water 8+12 (CPM)
< 1 CPM in normal surface lab condition
FOM H-3 in water 8+12 ml
> 600 without optimization
> 900 with optimization and Digital Pb
Dimensions, W/H/D (cm)
69/130/95
Weight (kg)
~800
Options
525-003 alpha/beta separation
462-019 External Eu-152 std source
425-2003 Cooling