X-AGM
Area Gamma Monitor
The X-AGM is an advanced gamma area monitor designed for continuous dose-rate detection, ideal for facilities requiring constant monitoring of gamma activity. It can function as a standalone unit or be integrated into a broader Environmental Monitoring System.
Its compact and clean design ensures no visible cables and minimizes dust accumulation, making it suitable for GMP installations in clean rooms. The device includes dual-microprocessor electronics, an integrated OLED display, and an energy-compensated GM tube for reliable radiation measurements
Additionally, the system supports external probes for flexible applications, such as monitoring in radiopharmaceutical preparation rooms or cyclotron vaults. It offers both standard and high-range dose rate probes to suit various monitoring needs.
Key Features
Versatile Installation
Dual-Microprocessor Electronics
Integrated OLED Display
Internal Energy-Compensated GM Tube
External Probe Option
Four Relay Outputs
RS-485 Communication Port
Visual and Audible Alarms
Two Configurable Alarm Thresholds
Password-Protected Software
Selectable Units of Measurement
Durable Design
GMP-Ready
Power Supply Included
Technical Specification
Detector
Halogen quenched GM tube.
Radiation Type
Gamma.
Energy Dependence
±15%.
Measurement Range
• Single Gamma Probe: 0.1 μSv/h to 10 mSv/h (0.01 mR/h to 100 mR/h).
• Dual Gamma Probe: 0.1 μSv/h to 1 Sv/h (0.01 mR/h to 100 R/h).
Accuracy
±15%.
High Voltage
Hardware-settable, typically 500V.
Display
128×64 white OLED matrix for long-lasting and high-visibility readouts.
Calibration
Factory calibration sheet provided as per ISO 9001 standards.
Response Time
2 seconds.
Alarm Indication
• Green LED: Normal operation.
• Yellow LED: Pre-alarm threshold exceeded.
• Red LED: Alarm threshold exceeded.
• Acoustic Signal: Activated when the alarm threshold is surpassed.
• Fault Condition: All LEDs are off, and a faulty indication is displayed on the OLED screen.
Controls
Alarm acknowledgment (muting acoustic signal), automatic restoration of normal conditions, high voltage control.
Outputs
Four settable relay outputs.
Communication Protocol
RS-485 with proprietary protocol.
Power Supply
110/230 VAC, 50/60 Hz.
Dimensions and Weight
125 mm diameter x 60 mm depth, 1 kg (without alarm column).
Operating Humidity Range
0% to 90% relative humidity.
Operating Temperature Range
0°C to 50°C.
Software Features
User-Friendly Interface
Simple and easy-to-use software with password protection for secure access.
Multi-Device Connection
Allows simultaneous connection of multiple devices.
Comprehensive Monitoring
Supports both wired and wireless communication with graphical displays and tailored report generation.
History Tracking
Maintains records of all events and alarms
Additional Options
External remote probe with a single energy-compensated GM tube.
External remote probe with dual energy-compensated GM tubes for high-range measurements, ideal for cyclotron and hot cell applications.
On-board or remote alarm tower with three lights.
Calibration certificate from an accredited laboratory.