X-Spec

Key Features

Digital Multi-Channel Analyze

Advanced digital MCA for performing spectrometry with scintillation detectors.

Compact and Complete

Includes all necessary components for spectrometry—pre-amplifier, high voltage power supply, digital processor, and MCA.

Wide Scintillator Support

Compatible with NaI(Tl), CsI, LaBr3(Ce), LaCl3(Ce), CeBr3, and HgI detectors using PMTs or SiPMs.

High Voltage Power Supply

Integrated HV power supply with low ripple, low electronic noise, and low power consumption.

Versatile Channels

MCA supports 256, 512, 1024, 2048, or 4096 channels.

USB or Ethernet Connectivity

Powered via USB or Power-over-Ethernet, ensuring easy integration with computers or networks.

Digital Signal Processing

Features settings for FlatTop, RiseTime, Digital Base-Line Restorer (BLR), and Digital Pile-Up Rejection.

Software Integration

Provided programming libraries compatible with MS Windows and Linux.

Two Acquisition Modes

Pulse-Height Analysis (PHA)

Technical Specification

PHA Acquisition Mode
Memory

256, 512, 1024, 2048, or 4096 channels.

Presets

Acquisition presets on live time, real time, or counts on a region of interest (ROI) or combination.

Discriminator Settings

Upper and lower level discriminator settings provided in channels.

Gain Settings
Amplification Factors

Gain settings with amplification factors of 1, 2, 4, and 8.

Fine Gain

Adjustable from 0.1 to 8.0 µsec in steps of 0.1 µsec.

Memory

256, 512, 1024, 2048, or 4096 channels.

Dwell Time

Ranges from 0.1 seconds to “count-forever.”

ROI Setup

Easily set up via stored ROIs or manually marked in PHA mode.

Digital Settings
Rise Time

Adjustable from 0.1 to 12 µsec in steps of 0.2 µsec.

Flat Top

Adjustable from 0.1 to 8.0 µsec in steps of 0.1 µsec.

Threshold

1 to 255.

Base Line Restoration

Digital BLR functionality.

Pile-Up Rejection

Integrated PUR system

High Voltage Power Supply
Voltage Range

0 to 1,500 V in 4096 steps.

Connection

USB 2.0 cable included.

Physical Specifications
Connections

USB Type Mini B and Ethernet.

Weight

120 grams.

Dimensions

Height 60 mm, Diameter 55 mm.

Temperature Range

Industrial range from -40°C to +85°C.

Indicators

Red LED for detector high voltage, Yellow LED for incoming count rate, and Green LED for power and communication status.

Software Features
Energy Calibration

Tools for peak search and energy calibration.

Basic Spectroscopy Software

Included software for data acquisition and visualization in PHA or MCS modes.

Spectrum Analysis

ROI-based spectrum analysis with the ability to save acquired spectra in ASCII format.

Report Generation

Generate detailed reports and maintain nuclide libraries.

External Application Support

Allows external applications to process spectra at the click of a button.